■Column
Oven:
Temperature
range: Room temperature + 2°C to 450°C (using liquid CO2 gas: -50°C to
450°C)
Temperature
program steps: Up to 32 (cooling program possible)
Injection
Units: Maximum 3 independently temperature controlled injector units
Injection
port unit: Split/splitless injection unit provided as standard.
◎Split/Splitless
Injection Unit (SPL-2030):Temperature range: room temperature + 5°C to 450°C
◎Direct
Injection Unit (WBI-2030)
◎On-column/Programmable
Temperature Vaporizer Injection Unit (OCI/PTV-2030)
■
Detectors:
Four
detectors can be installed simultaneously and individually temperature
controlled.
◎Flame
Ionization Detector (FID):Temperature range: to 450°C
Minimum detected quantity: 1.2 pg C/s (dodecane)
◎Thermal
Conductivity Detector (TCD):Temperature range: to 400°C
Sensitivity: 20000 mV · mL/mg (decane)
◎Electron
Capture Detector (ECD):Temperature range: to 400°C
Minimum detected quantity: 4 fg/s (lindane)
◎Flame
Photometric Detector (FPD): Temperature range: to 450°C
Minimum detected quantity: P 45 fg P/s (tributyl phosphate) S 2 pgS/s
(dodecane thiol)
◎Flame
Thermionic Detector (FTD, also called NPD or TSD):Temperature range: to 450°C
Minimum detected quantity: N 0.1 pg N/s (azobenzene) P 0.01 pg P/s
(malathion)
◎Barrier
Discharge Ionization Detector (BID): Temperature range: to 350°C
Minimum detected quantity: 0.8 pg C/s (dodecane)
■ Flow
Control Unit
■
Split/splitless mode:
Pressure
setting range: 0 to 1035 kPa
Programmable
steps: 7 (pressure program possible)